Part Number |
Description |
UK 240V |
EUR 220V |
US 110V |
K0UK6075M001 |
K0006075M001 |
K0US6075M001 |
SP60 Portable Sphere Spectrophotometer 8mm Fixed Aperture |
Measuring Geometrics |
d/8°, DRS spectral engine, fixed aperture: 8mm viewing/12mm illumination |
Light Source |
Gas filled tungsten lamp |
Illuminant Types |
C, D50, D65, D75, A, F2, F7, F11, F12 |
Standard Observers |
2° and 10° |
Spectral Range |
400-700nm |
Memory |
1,024 standards with tolerances, 2,000 samples |
Measurement Range |
0 to 200% reflectance |
Measuring Time |
Approximately 2 seconds |
Inter-Instrument Agreement |
CIE L*a*b*:
Average 0.40 ΔE*ab based on average of 12 BCRA Series II tiles
(specular component included)
Maximum 0.60 ΔE*ab on any tile (specular component included)
CMC Equivalent:
Average 0.30 ΔEcmc based on average of 12 BCRA Series II tiles
(specular component included)
Maximum 0.50 ΔEcmc on any tile (specular component included) |
Short-term Repeatability* |
0.10 ΔE*ab on white ceramic (standard deviation) |
Lamp Life |
Approximately 500,000 measurements |
Power Supply |
Removable (Ni-metal hydride) battery pack; 7.2DC rated @1450mAh |
Measurements per Charge |
1000 measurements within 8 hour period |
Weight |
1.1kg (2.4lb) |
Dimensions |
109 x 83 x 195mm (4.3 x 3.3 x 7.7”) |
Certificate Available |
Certificate of Calibration: issued for Fixed Calibration equipment and shows readings and traceability. |
Packing List |
Elcometer 6075/1, calibration standards, AC adaptor, carry case & operating instructions |
Only available for sales within the EU
*Based on 20 measurements on a white tile
Can be used in accordance with:
AS/NZS 1580.601.3, ASTM C 609, ASTM D 2244, ASTM E 1164, ASTM E 308, ASTM E 313, BS 8493, DIN 5033-2,
DIN 5033-3, DIN 5033-4, DIN 5033-7, DIN 6174, EN 12373-12, EN 13523-15, ISO 7724-2, ISO 7724-3, JIS K 5600-4-5, JIS K 5600-4-6, NF T36-006,
NF X08-012-1, NF X08-012-2